Price:   $682
Cat Number:  E-CL-H0050
Lead Time:   Order now, ship in 3 days

Reactivity:  Human

Detection Range:  7.81~500 pg/mL

Sensitivity:  4.69 pg/mL

Add to cart

Intended use

This CLIA kit applies to the in vitro quantitative determination of Human IP-10/CXCL10 concentrations in serum, plasma and other biological fluids.

Test principle

This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human IP-10/CXCL10. Standards or samples are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human IP-10/CXCL10 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IP-10/CXCL10, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Human IP-10/CXCL10. You can calculate the concentration of Human IP-10/CXCL10 in the samples by comparing the RLU value of the samples to the standard curve.

Assay type Sandwich-CLIA
Format 96T
Assay time 3.5h
Reactivity Human
Detection method Chemiluminescence
Detection range 7.81-500 pg/mL
Sensitivity 4.69 pg/mL
Sample volume 100μL
Sample type serum, plasma and other biological fluids
Repeatability CV < 15%

Specificity

This kit recognizes Human IP-10/CXCL10 in samples. No significant cross-reactivity or interference between Human IP-10/CXCL10 and analogues was observed.

Typical data

As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

Concentration
(pg/mL)
RLU Average Corrected
500 46893
56181
51537 51508
250 19287
22709
20998 20969
125 9949
8843
9396 9367
62.5 4279
4745
4512 4483
31.25 2425
2173
2299 2270
15.63 1350
1150
1250 1221
7.81 716
764
740 711
0 28
30
29 --
CLIA-Elabscience

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human IP-10/CXCL10 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human IP-10/CXCL10 were tested on 3 different plates, 20 replicates in each plate.

  Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/mL) 24.60 72.26 192.19 24.27 73.23 175.93
Standard deviation 2.01 6.56 15.34 2.60 8.41 17.43
C V (%) 8.17 9.08 7.98 10.71 11.48 9.91

Recovery

The recovery of Human IP-10/CXCL10 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range (%) Average Recovery (%)
Serum (n=5) 94-111 102
EDTA plasma (n=5) 100-115 107
Cell culture media (n=5) 85-96 91

Linearity

Samples were spiked with high concentrations of Human IP-10/CXCL10 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

    Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 93-106 92-105 102-118
Average (%) 98 98 109
1:4 Range (%) 84-96 101-116 98-115
Average (%) 91 109 105
1:8 Range (%) 96-110 87-101 97-111
Average (%) 103 94 104
1:16 Range (%) 103-116 101-117 93-105
Average (%) 109 109 98

Kit Components& Storage

An unopened kit can be stored at 4℃ for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.

Item Specifications Storage
Micro CLIA Plate(Dismountable) 8 wells ×12 strips -20℃, 6 months
Reference Standard 2 vials
Concentrated Biotinylated Detection Ab (100×) 1 vial, 120 μL
Concentrated HRP Conjugate (100×) 1 vial, 120 μL -20℃(shading light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 4℃, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL
HRP Conjugate Diluent 1 vial, 14 mL
Concentrated Wash Buffer (25×) 1 vial, 30 mL
Substrate Reagent A 1 vial, 5 mL 4℃ (shading light)
Substrate Reagent B 1 vial, 5 mL 4℃ (shading light)
Plate Sealer 5 pieces
Product Description 1 copy
Certificate of Analysis 1 copy

Note: All reagent bottle caps must be tightened to prevent evaporation and microbial pollution.
The volume of reagents in partial shipments is a little more than the volume marked on the label, please use accurate measuring equipment instead of directly pouring into the vial(s).

Other supplies required

  • Chemiluminescence immunoassay analyzer
  • High-precision transfer pipette, EP tubes and disposable pipette tips
  • Incubator capable of maintaining 37℃
  • Deionized or distilled water
  • Absorbent paper
  • Loading slot for Wash Buffer

Assay procedure

1.Add 100 μL of standard or sample to each well. Incubate for 90 min at 37℃.

2.Remove the liquid.

3.Add 100 μL Biotinylated Detection Ab. Incubate 1 hour at 37℃. Aspirate and wash 3 times.

4.Add 100 μL HRP Conjugate. Incubate 30 min at 37℃. Aspirate and wash 5 times.

5.Add 100 μL Substrate Mixture Solution. Incubate for 5 min at 37℃.

6.Fluorescence appeared. Measure the RLU value with the Chemiluminescence immunoassay analyzer. Calculation of the results.

If you have any question about the products, please contact us by clicking the message button near the home button of your phone, we will reply you in 12 hours.