Mouse IP-10/CXCL10 (Interferon Gamma Induced Protein 10kDa) CLIA Kit

Size:  
Qty:
- +
Price:   $682
Cat Number:  E-CL-M0020
Lead Time:   Order now, ship in 3 days

Reactivity:  Mouse

Detection Range:  12.50~800 pg/mL

Sensitivity:  7.50 pg/mL

Add to cart

Intended use

This CLIA kit applies to the in vitro quantitative determination of Mouse IP-10/CXCL10 concentrations in serum, plasma and other biological fluids.

Test principle

This CLIA kit uses the Sandwich-CLIA principle. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse IP-10/CXCL10. Standards or samples are added to the micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse IP-10/CXCL10 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse IP-10/CXCL10, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Mouse IP-10/CXCL10. You can calculate the concentration of Mouse IP-10/CXCL10 in the samples by comparing the RLU value of the samples to the standard curve.

Assay type Sandwich-CLIA
Format 96T
Assay time 3.5h
Reactivity Mouse
Detection method Chemiluminescence
Detection range 12.50-800 pg/mL
Sensitivity 7.50 pg/mL
Sample volume 100μL
Sample type serum, plasma and other biological fluids
Repeatability CV < 15%

Specificity

This kit recognizes Mouse IP-10/CXCL10 in samples. No significant cross-reactivity or interference between Mouse IP-10/CXCL10 and analogues was observed.

Typical data

As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.

Concentration
(pg/mL)
RLU Average Corrected
800 51825
53459
52642 52616
400 19859
23609
21734 21708
200 10041
9375
9708 9682
100 4193
4911
4552 4526
50 2289
2087
2188 2162
25 1094
1026
1060 1034
12.50 489
529
509 483
0 26
26
26 --
CLIA-Elabscience

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse IP-10/CXCL10 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse IP-10/CXCL10 were tested on 3 different plates, 20 replicates in each plate.

  Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/mL) 43.41 114.92 324.95 46.74 119.84 304.31
Standard deviation 5.35 12.03 26.97 3.78 10.46 19.02
C V (%) 12.32 10.47 8.30 8.09 8.73 6.25

Recovery

The recovery of Mouse IP-10/CXCL10 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range (%) Average Recovery (%)
Serum (n=5) 95-108 100
EDTA plasma (n=5) 88-101 95
Cell culture media (n=5) 90-102 96

Linearity

Samples were spiked with high concentrations of Mouse IP-10/CXCL10 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

    Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 96-106 98-113 91-105
Average (%) 101 106 97
1:4 Range (%) 100-118 91-103 91-103
Average (%) 107 96 97
1:8 Range (%) 94-109 100-118 103-116
Average (%) 101 107 110
1:16 Range (%) 99-112 90-103 89-100
Average (%) 106 95 94

Kit Components& Storage

An unopened kit can be stored at 4℃ for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.

Item Specifications Storage
Micro CLIA Plate(Dismountable) 8 wells ×12 strips -20℃, 6 months
Reference Standard 2 vials
Concentrated Biotinylated Detection Ab (100×) 1 vial, 120 μL
Concentrated HRP Conjugate (100×) 1 vial, 120 μL -20℃(shading light), 6 months
Reference Standard & Sample Diluent 1 vial, 20 mL 4℃, 6 months
Biotinylated Detection Ab Diluent 1 vial, 14 mL
HRP Conjugate Diluent 1 vial, 14 mL
Concentrated Wash Buffer (25×) 1 vial, 30 mL
Substrate Reagent A 1 vial, 5 mL 4℃ (shading light)
Substrate Reagent B 1 vial, 5 mL 4℃ (shading light)
Plate Sealer 5 pieces
Product Description 1 copy
Certificate of Analysis 1 copy

Note: All reagent bottle caps must be tightened to prevent evaporation and microbial pollution.
The volume of reagents in partial shipments is a little more than the volume marked on the label, please use accurate measuring equipment instead of directly pouring into the vial(s).

Other supplies required

  • Chemiluminescence immunoassay analyzer
  • High-precision transfer pipette, EP tubes and disposable pipette tips
  • Incubator capable of maintaining 37℃
  • Deionized or distilled water
  • Absorbent paper
  • Loading slot for Wash Buffer

Assay procedure

1.Add 100 μL of standard or sample to each well. Incubate for 90 min at 37℃.

2.Remove the liquid.

3.Add 100 μL Biotinylated Detection Ab. Incubate 1 hour at 37℃. Aspirate and wash 3 times.

4.Add 100 μL HRP Conjugate. Incubate 30 min at 37℃. Aspirate and wash 5 times.

5.Add 100 μL Substrate Mixture Solution. Incubate for 5 min at 37℃.

6.Fluorescence appeared. Measure the RLU value with the Chemiluminescence immunoassay analyzer. Calculation of the results.

If you have any question about the products, please contact us by clicking the message button near the home button of your phone, we will reply you in 12 hours.