Product Details

Properties

Assay type Sandwich-ELISA
Format 96T/48T/24
Assay time 1.5h
Detection range 0.78-50 ng/mL
Sensitivity 0.47 ng/mL
Sample type &Sample volume serum and plasma; 50μL
Specificity This kit recognizes CEA in samples. No significant cross-reactivity or interference between CEA and analogues was observed.
Reproducibility Both intra-CV and inter-CV are < 10%.
Application This ELISA kit applies to the in vitro quantitative determination of CEA concentrations in serum and plasma.

Test Principle

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human CEA. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human CEA are added to the micro ELISA plate wells. Human CEA in samples (or standards) combines with the coated antibody and HRP linked detection antibody specific to CEA. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Human CEA in the samples is then determined by comparing the OD of the samples to the standard curve.

Kit components & Storage

An unopened kit can be stored at 2-8℃ for 6 months. After test, the unused wells and reagents should be stored according to the table.
ItemSpecificationsStorage
Micro ELISA Plate(Dismountable) 96T: 8 wells ×12 strips
48T: 8 wells ×6 strips
24T: 8 wells ×3 strips
96T*5: 5 plates, 96T
2-8℃, 1 months
Reference Standard 96T: 2 vials
48T/24T: 1 vial
96T*5: 10 vials
2-8℃, use the reconstituted standard within 24h
Concentrated HRP Linked Detection Ab(100×) 96T: 1 vial, 60 μL
48T/24T: 1 vial, 30 μL
96T*5: 5 vials, 60 μL
2-8℃(Protect from light)
Reference Standard & Sample Diluent 96T/48T/24T: 1 vial, 20 mL
96T*5: 5 vials, 20 mL
2-8℃
HRP Linked Ab Diluent 96T/48T/24T: 1 vial, 14 mL
96T*5: 5 vials, 14 mL
Concentrated Wash Buffer(25×) 96T/48T/24T: 1 vial, 30 mL
96T*5: 5 vials, 30 mL
Substrate Reagent 96T/48T/24T: 1 vial, 10 mL
96T*5: 5 vials, 10 mL
2-8℃(Protect from light)
Stop Solution 96T/48T/24T: 1 vial, 10 mL
96T*5: 5 vials, 10 mL
2-8℃
Plate Sealer 96T/48T/24T: 5 pieces
96T*5: 25 pieces
Product Description 1 copy
Certificate of Analysis 1 copy

Technical Data

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level CEA were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level CEA were tested on 3 different plates, 20 replicates in each plate.

  Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (ng/mL) 2.44 7.17 18.89 2.29 6.53 19.86
Standard deviation 0.13 0.40 0.60 0.13 0.26 0.66
CV (%) 5.16 5.51 3.16 5.79 4.01 3.32

Recovery

The recovery of CEA spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range (%) Average Recovery (%)
Serum(n=8) 93-107 100
EDTA plasma(n=8) 87-98 92

Linearity

Samples were spiked with high concentrations of Human CEA and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Target Information

Database LinksSwissProt: P06731
SynonymsCEA,Carcinoembryonic Antigen
Research AreaCancer, Immunology, Stem cells, Tags and Cell Markers

Assay Procedures

If you have any question about the products, please contact us by clicking the message button near the home button of your phone, we will reply you in 12 hours.